8-5
Message
Alarm
Priority
Cause
Solution
calibration.
4. Replace the VCM board and perform
calibration.
Calibrate Flow
Sensor and Insp
Valve
Low
1. The calibration table is not
found in the EEPROM.
2. The checksum of
Calibration table does not
match.
Perform maintenance calibration. Refer
to chapters related to the flow sensor
calibration.
Calibrate
Pressure Sensor
and PEEP Valve
Low
1. The calibration table is not
found in the EEPROM.
2. The checksum of
Calibration table does not
match..
Perform maintenance calibration. Refer
to chapters related to the pressure
sensor calibration.
Calibrate O2
Sensor
Low
1. The calibration table is not
found in the EEPROM.
2. The checksum of
Calibration table does not
match.
3. The AD difference between
the 100% and 21% O2
concentrations in the O2
sensor table is smaller than
79AD.
1. Calibrate the O2 sensor.
2. If the problem persists, replace the
O2 sensor.
Ventilator
Initialization
Error
Ventilator
Initialization:
Time out
High
After the machine is powered
on, the CPU board cannot send
the control parameter settings
to the VCM board.
1. Restart the anaesthesia machine.
2. Reconnect or replace the
communication line between the CPU
board and the VCM board.
3. If the problem persists, replace the
VCM board.
4. If the problem persists, replace the
CPU board.
Drive Gas
Pressure Low
High
The drive gas pressure is low.
1. Check whether the gas supply is
normal.
2. Short circuit the pressure switch.
The alarm should disappear. If the
alarm does not disappear, the pressure
switch fails. Replace the pressure
switch. If the pressure switch does not
fail, check the connection between the
pressure switch and the VCM board.
3. If both of the preceding items are
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...