4-19
If any damage or loose connection is detected, replace the corresponding part.
4.10.6 Negative Pressure Suction Inspection
4.10.6.1 Check the Tube Connections of Liquid Collection Bottle
1. Check if the tubes are correctly connected following the connection diagram printed on
the liquid collection bottle. If not, the overfill protection device cannot work normally.
2. Check if the tubes are inserted in place. If not, leakage may occur, resulting in failure to
reach negative pressure of 50 kPa.
4.10.6.2 Check the Negative Pressure
1. Read the reading on the Air pipeline pressure gauge on the front side of the anesthesia
machine and make sure that the pipeline supply gas pressure is within the normal
pressure range. (Only for the anesthesia machines without Venturi negative pressure
suction configuration.)
2. Set the swapping switch on the negative pressure regulator to REG position.
3. Set the switch of Venturi negative pressure generator to ON position. (Only for the
anesthesia machines without Venturi negative pressure suction configuration.)
4. Occlude the patient connection outlet of suction tube with hand.
5. Check if the reading on the negative pressure regulator is greater than 50 kPa. If it is less
than 50 kPa, check the following:
a. Increase the pressure at the rear end of negative pressure regulator clockwise.
b. Rotate counterclockwise to remove the muffler. Shake dust and foreign substance
from the muffler.
Sealing ring
Hose and connector
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...