8-24
Message
Alarm
Priority
Cause
Solution
CO2 module.
EtCO2
Overrange
Low
The monitored value is
out of the measurable
range.
1. Re-calibrate the CO2 module.
2. Replace the CO2 module.
FiCO2
Overrange
Low
The monitored value is
out of the measurable
range.
1. Re-calibrate the CO2 module.
2. Replace the CO2 module.
CO2 Zero
Failed
Low
The CO2 module is
faulty.
1. Re-zero the CO2 module.
2. Replace the CO2 module.
CO2 Init Error
High
CO2 initialization error
Replace the CO2 module.
Incompatible
CO2 Software
Version
High
The CO2 module is not
compatible with the
software version.
Replace the CO2 module (the M02A
module is not supported).
8.2.9 Mainstream CO
2
module:
Message
Alarm
Priority
Cause
Solution
EtCO2
Overrange
Low
The monitored value is
out of the measurable
range.
1. Re-calibrate the CO2 module.
2. Replace the CO2 module.
FiCO2
Overrange
Low
The monitored value is
out of the measurable
range.
1. Re-calibrate the CO2 module.
2. Replace the CO2 module.
CO
2
Zero
Failed
Low
The CO2 module is
faulty.
1. Re-zero the CO2 module.
2. Replace the CO2 module.
CO2 Init Error
High
CO2 initialization error
Replace the CO2 module.
CO2 Sensor
Error
Low
The sensor of the
mainstream CO2 module
is faulty.
Replace the CO2 module.
CO2 No
Sensor
Low
The sensor of the
mainstream CO2 module
is not connected, or a
communication error
occurs.
Replace the CO2 module.
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...