8-4
Message
Alarm
Priority
Cause
Solution
and expiratory valve assembly.
4. When necessary, replace the VCM
board.
5. When necessary, replace the
expiratory valve assembly..
6. When necessary, replace the power
board.
Insp Valve
Failure
Medium
1. Inspiratory valve voltage
error2. Inspiratory valve flow
error
1. Check whether the accuracy of the
inspiratory flow sensor is within the
valid range. On the [Maintenance]
menu, calibrate the flow sensor. Or,
when a flow error occurs, replace the
flow sensor.
2. Test the voltage at the corresponding
testing point.
3. Check the installation of and
connection between the power cable
and expiratory valve assembly.
4. When necessary, replace the VCM
board.
5. When necessary, replace the
expiratory valve assembly..
6. When necessary, replace the power
board.
Safety Valve
Failure
Medium
PEEP safety valve voltage
error
1. Test the voltage at the corresponding
testing point.
2. Check the installation of and
connection between the power cable
and expiratory valve assembly.
3. When necessary, replace the VCM
board.
4. When necessary, replace the
expiratory valve assembly.
5. When necessary, replace the power
board.
Flow Sensor
Failure
Low
The ventilator flow is out of
range.
1. Check whether the zero point of the
flow sensor is within the valid range.
2. Check whether the monitored value
of the flow sensor is within the valid
range.
3. Replace the flow sensor and perform
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...