4-30
4.12.2 Test the O
2
Concentration Monitoring and Alarms
NOTE
This test is not required if O
2
sensor is not configured.
1. Set the Bag/vent switch to the
position.
2. Remove the O
2
sensor and make sure that the sensor measures approximately 21% O
2
in
room air.
3. Select
[
Alarm
] and then [
Limit
]. Set the FiO
2
low limit to 50%.
4. Make sure that a low FiO
2
alarm occurs.
5. Set the FiO
2
low limit to a value less than the measured FiO
2
value and make sure that
the alarm cancels.
6. Put the O
2
sensor back in the circuit.
7. Select
[
Alarm
] and then [
Limit
]. Set the FiO
2
high limit to 50%.
8. Connect the manual bag to the manual bag port. Push the O
2
flush button
to fill the
manual bag. After two to three minutes, make sure that the sensor measures
approximately 100% O
2.
9. Make sure that a high FiO
2
alarm occurs.
10. Set the FiO
2
high limit to 100% and make sure that the alarm cancels.
4.12.3 Test the Low Minute Volume (MV) Alarm
1. Make sure that MV alarm is turned on.
2. Select
[
Alarm
] and then [
Limit
]. Set the MV low limit to 8.0 L/min.
3. Make sure that a low MV alarm occurs.
4. Select
[
Alarm
] and then [
Limit
]. Set the MV low alarm limit to the default.
4.12.4 Test the Apnea Alarm
1. Connect the manual bag to the manual bag port.
2. Set the Bag/vent switch to the
position.
3. Turn the APL valve control to set the APL valve to the minimum position.
4. Inflate the manual bag to make sure that a complete breathing cycle occurs.
5. Stop inflating the manual bag and wait for at least 20 seconds to make sure that the
apnea alarm occurs.
6. Inflate the manual bag to make sure that the alarm cancels.
4.12.5 Test the Sustained Airway Pressure Alarm
1. Connect the manual bag to the manual bag port.
2. Turn the O
2
flow control to set the O
2
flow to minimum.
3. Turn the APL valve control to set the APL valve to 30 cmH
2
O position.
4. Set the Bag/vent switch to the
position.
5. Push the O
2
flush button for approximately 15 seconds. Make sure that the sustained
airway pressure alarm occurs.
6. Open the patient connection and make sure that the alarm cancels.
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...