9-16
9.2.15 Replace Fuse
1. Remove the covers of fuses by flathead screwdriver according to the direction on the
bases of fuse.
2. Insert new fuses, and tighten the covers of fuses by flathead screwdriver.
WARNING
Bafore replace fuse, ensure to turn off the system switch and disconnect power
cable to avoid electrical shock.
Ensure new fuse meets the requirements of the spacifications on the Operator’s
manual of this product.
9.2.16 Replace Battery
1. Open the service door.
2. Unscrew the three (3) screws fixing the cover plate of battery box, to remove lithium
battery.