3-19
1. Align the vaporizer over the valve cartridges of the mounting bar. Hang the vaporizer on
the mounting bar as shown below. Note that the locking handle is in the unlocked
position.
2. Rotate the locking handle clockwise to the locked position as shown below.
3.2.4.1 Assemble the Vaporizer
1. Mount the vaporizer onto the manifold.
2. Push and turn the locking lever clockwise to lock the vaporizer in position.
3. Ensure that the top of the vaporizer is horizontal. If not, remove the vaporizer and
reinstall it.
Locking handle in
the unlocked
position
Locking handle in the
locked position