7-31
7. Connect the 4-way assembly to the pressure sensor P1, PEEP pressure sensor P2, and
#97 sampling line on the VCM, and the low-pressure port on the Fluke VT Plus (or
high-pressure port sampling interface on the VT, or sampling interface on the FPM).
The #99 sampling line is not connected to the assembly during this calibration.
Airway connection involving the calibration device VT Plus
High-pressure port sampling interface on the calibration device VT
#97 sampleline
Connecting to the
low-pressure port on
the Fluke VT Plus
(or high-pressure
port on the VT)
Pressure sensor
P1
Pressure sensor
P2
#99
sampleline