6-22
6.6.5 Check the Flow Sensor Accuracy
NOTE
If a great deviation of TV measured value occurs, test the measurement accuracy
of flow sensors so as to determine whether to perform flow calibration again.
Checking the flow sensor accuracy is to check:
The measurement consistency between the flow sensors inside the machine;
The measurement accuracy of any single flow sensor inside the machine;
The validity of (factory) flow calibration result.
To check the measurement accuracy of flow sensors:
1. Make sure that the circuit, calibration device (or other flow measurement device) and
breathing tubes are connected in serial, similar to tubes connection in flow calibration.
For details, refer to
7.3.2
Flow calibration (Service)
.
2. When the system is Standby, select the [
Maintenance
] shortcut key
→
[
Factory
Maintenance
>>]
→
[
Diagnostic Test
>>]
→
[
Valves—Test Tool
>>] to access the
[
Valves—Test Tool
] menu.
3. Set PEEP safety valve to [
ON
].
4. Set the D/A value of the PEEP valve to above 1500 and ensure that the pressure at
which the expiratory valve closes is above 30 cmH2O.
5. Increase the D/A value of the inspiratory valve, causing the measured flow value of the
anesthesia machine calibration device to fall with the following ranges respectively:
(3±0.5) L/min, (10±1) L/min, (20±1) L/min, (30±2) L/min, (60±3) L/min. Record the
measured flow values of the inspiratory flow sensor, expiratory flow sensor and
ventilator flow sensor to which each setting corresponds respectively. When testing the
flow accuracy of (60±3)L/min, only compare the accuracy of the inspiratory flow sensor
and expiratory flow sensor. It is no need to compare the ventilator flow sensor.
6. Make sure that the deviation between the measured data of the inspiratory flow sensor,
expiratory flow sensor and ventilator flow sensor and that of the anesthesia machine
calibration device must not exceed 1 L/min or 5% of the measured value of the
calibration device, whichever is greater. Otherwise, refer to
7.3.2
Flow calibration
(Service)
to perform flow calibration again.
7. If anesthesia machine calibration device is unavailable, you can execute steps 1 through
5 to test the accuracy of flow sensors. The deviation between the measured data of the
inspiratory flow sensor and expiratory flow sensor and that of the ventilator flow sensor
must not exceed 1 L/min or 5% of the measured value of the ventilator flow sensor,
whichever is greater. Otherwise, refer to
7.3.2
Flow calibration (Service)
to perform
flow calibration again.
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...