9.2.36 Remove Drive Gas Switching Valve
1. Remove the rear cover plate and cart rear cover plate.
2. Disconnect the cables and pipeline of the drive gas switching valve.
3. Unscrew the two (2) screws fixing the drive gas switching valve.
9.2.37 Remove Power Cable
Unscrew the three (3) screws on the anti-pulling hook of mains plug, and then remove the power
cable.