3-12
3. Loosen the five screws fixing the upper cover of the air compressor from behind, pull
out the upper cover assembly of the air compressor from the front, remove the cable
materials and pipelines connected to the upper cover, and take out the upper cover.
4. Use an inner hexagon screw driver to loosen the two transport protection screws fixing
the air pump.
Inner hexagon screw driver
Transport protection
screw