3-18
3.2.3.1 Cylinder Wrench and User Guide
1. Fix the cylinder wrench onto the back of this device so that it can be used to open or
close cylinders without disconnecting from the machine.
2. Hang the user guide on the device handle.
3.2.4 Vaporizer
WARNING
If the vaporizer is incompatible with this device Anesthesia System, the vaporizer
will not work at all. Use vaporizers with Selectatec mounting system that are
compliant to ISO8835-4. Refer to the vaporizer manufacturer’s Instructions For
Use for filling or draining the vaporizer and other information.
For this device Anesthesia System, using or turning on more than one vaporizer
simultaneously is mechanically prevented by the Selectatec mount system. Do not
attempt to override this safety feature.
Use care in lifting and manipulating vaporizers during the mounting process as
their weight may be greater than expected, based on their size and shape.
NOTE
The barometric pressure at the customer side may be different from the factory
calibration pressure of anesthetic vaporizer. This may cause inaccurate
concentration output of anesthetic agent. The operator should continuously
monitor the concentration of anesthetic agent during system use.
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...