8-72
Failure
Description
Possible Cause
Recommended Action
The inlet flow sensor on the
integrated airway of the
expiratory valve is damaged.
Replace the expiratory valve or its integrated
airway.
Expiration begins in advance
because the specified Plimit
value is too small.
Set the Plimit to a larger value so that the
Paw value does not exceed the limit.
The displayed TVe and TVi
values are different.
In the valve test tool, compare the
measurement errors of the three sensors and
determine whether to calibrate the sensors
according to section
7.3.2 Flow calibration
(Service)
.
In the preceding table, the items marked with an asterisk (*) are possible causes of inaccurate
measurement values of the flow sensor. Perform the following steps to check whether the fault of
inaccurate breathing capacity is induced by the causes marked with an asterisk (*):
1. Turn
off
the
flow regulator.
2. Ensure that the patient is disconnected from the system and the auto/manual ventilation
switch is turned to the mechanical ventilation position.
3. Remove the bellows and then install the bellow housing correctly.
4. Remove the water collection cup.
5. Connect the inspiratory connector and the expiratory connector by using a corrugated
tube. See the figure below.
The water collection cup
has been removed.
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...