8.4.1.2 Precautions for Use of Test Fixture of Vaporizer Support
Before use, take out the seal ring on the connector of the vaporizer support for cooperating with
the vaporizer. Slide the fixture into the connector, as shown in the following figures.
Rotate the pressure head of the test fixture of the vaporizer support clockwise to the end until the
lower surface of the test fixture of the vaporizer support is in contact with the upper surface of the
vaporizer support connector, as shown in the following figure.
After the test fixture slid in
Before the test
fixture slid in