4.12 Alarm Tests
4.12.1 Prepare for Alarm Tests
1. Connect a test lung or manual bag to the Y piece patient connection.
2. Set the Bag/vent switch to the
position.
3. Set the system switch to the
position.
4. Set the system to Standby.
5. Set the ventilator controls as follows:
Ventilation mode: select [
Vent Mode
] and then [
VCV
].
[
TV
]: 500 ml.
[
Rate
]:12 BPM.
[
I:E
]:
1:2.
[
Plimit
]: 30 cmH
2
O.
[
PEEP
]: OFF.
[
Tpause
]: OFF.
6. Push the O
2
flush button to fill the bellows, which rises to the top.
7. Select single flow control mode of fresh gas in the main menu.
8. Turn the O
2
flow control to set the O
2
flow to 0.5 to 1 L/min.
9
Press the soft key and select [
Ok
] from the pop-up menu to exit Standby.
10. Make sure that:
The ventilator displays the correct data.
The bellows inflates and deflates normally during mechanical ventilation.
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...