8-19
Message
Alarm
Priority
Cause
Solution
the module.
3. Replace the AG module.
ISO Over
Range
Low
The monitored value
exceeds the
measurable range of
the module.
1. Reduce the concentration of the monitored
gas to the normal range.
2. Calibrate the AG module again.
3. Replace the AG module.
SEV Over
Range
Low
The monitored value
exceeds the
measurable range of
the module.
1. Reduce the concentration of the monitored
gas to the normal range.
2. Calibrate the AG module again.
3. Replace the AG module.
DES Over
Range
Low
The monitored value
exceeds the
measurable range of
the module.
1. Reduce the concentration of the monitored
gas to the normal range.
2. Calibrate the AG module again.
3. Replace the AG module.
Rate Over
Range
Low
The monitored value
exceeds the
measurable range of
the module.
1. Reduce the respiration rate to the normal
range.
2. Calibrate the AG module again.
3. Replace the AG module.
8.2.7 BIS Module Runtime Alarms
Message
Alarm
Priority
Cause
Solution
BIS Init Error
High
The system cannot
receive the initialization
information of the BIS
module.
1. Reconnect the BIS module.
2. If the problem persists, replace
the BIS module.
BIS Comm
Abnormal
High
The waveform data
packets are not received
in 2s consecutively, or
data packets of
monitoring parameters of
the BIS module are not
received in 4s
consecutively.
1. Check the communication cable
between the BIS module and the
main system.
2. If the problem persists, replace
the BIS module.
BIS Over
Range
Low
The monitored value
exceeds the measurable
range of the module.
Replace the BIS module.
BIS L Over
Range
Low
The monitored value of
BIS L exceeds the
measurable range.
Replace the BIS module.
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...