8-16
Message
Alarm
Priority
Cause
Solution
Channel
Failure
PEEP sensor or Paw
sensor is out of range.
is out of range, replace the ventilation control
board or VCM board assembly.
2. If the AD of the zero point of each sensor is
within the valid range, check whether the
software of the VCM board is compatible with
the system software version. If incompatible,
upgrade the software of the VCM board to the
compatible version. If compatible, perform the
factory pressure calibration. If the problem
persists after the calibration, replace the
ventilator control board or VCM board
assembly.
Aux Control
Module
Voltage
The VPM 1.3 V
supply voltage is not
within the range of
1.121 V to 1.516 V.
1. Replace the ventilator control board or VCM
board assembly.
Pressure
Monitoring
Channel
Failure
For VCM:
1. The monitored
value of PEEP sensor
or Paw sensor is out of
range.
2. The zero point of
the PEEP sensor or
Paw sensor is
abnormal.
3. The PEEP sensor is
reversely connected.
1. Check whether the pressure sampling line is
connected correctly.
2. Zero the pressure and flow of each sensor.
3. Check whether the AD of the zero point of
each sensor is within the valid range. If the AS
is out of range, replace the ventilator control
board or VCM board assembly.
4. If the AD of the zero point of each sensor is
within the valid range, check whether the
software of the VCM board is compatible with
the system software version. If incompatible,
upgrade the software of the VCM board to the
compatible version. If compatible, perform the
factory pressure calibration. If the problem
persists after the calibration, replace the
ventilator control board or VCM board
assembly.
Keyboard Runtime Alarms
Message
Alarm
Priority
Cause
Solution
Key Error
Medium
A hard key is pressed
for more than 35s.
Check whether any hard key is continuously
pressed.
Keyboard
Comm Stop
Medium
1. The CPU board
cannot communicate
1. Restart the anaesthesia machine.
2. Check the cable connection between the
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...