8-50
Test procedures:
1. Turn off the pipeline gas supplies and switch on the O2 flush switch to release the
remaining pressure.
2. Remove the PU tube 83 connecting to the auxiliary oxygen flowmeter assembly and seal
this end with the tube plug.
3. Remove the PU tube 84 connecting to the auxiliary oxygen flowmeter assembly.
Connect the negative pressure ball and then squeeze the negative pressure ball to
exhaust the internal gas.
4. Release the negative pressure ball. If the negative pressure ball is completed expanded
within 30s, the auxiliary oxygen flowmeter assembly is damaged.
8.4.3.3 Leak Test for the Backup Oxygen Flowmeter Assembly
Tools:
Negative pressure ball (quantity: 1)
3106-06-00 adapter connector (quantity: 1)
3126-06-00 tube plug (quantity: 1)
PU tube (6 x 100) (quantity: 1)
Test procedures:
1. Turn off the pipeline gas supplies and switch on the O2 flush switch to release the
remaining pressure.
2. Remove the PU tube 45 connecting to the backup oxygen flowmeter assembly and seal
this end with the tube plug.
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...