7-21
12. The interface as shown in the following figure is displayed after the calibration is complete.
If the flow sensor calibration fails, the interface as shown in the following figure is
displayed. Read the detailed information displayed on the calibration device,
troubleshoot the failure and rectify the fault. Select [
Retry
] to re-calibrate. Select [
OK
]
to quit calibration.
If the flow sensor calibration succeeds, the interface as shown in the following figure is
displayed. Select [
OK
] to quit calibration.