7-30
4. Perform automatic calibration according to Step 4 in section 4.3.2.1. Connect the
calibration device and the anesthesia machine with a communication cable.
5. Press
[
Next
]. The menu as shown in the following figure is displayed:
Connect the sampling lines for pressure calibration with a 4-way assembly. The
following figure shows the 4-way assembly, calibration device connector, and VCM
used for pressure calibration.
6. Remove the #97 and #99 sampling lines from the pressure sensor. (See the following
figure.)
4-way assembly
connecting
pressure
calibration
samplelines