2-52
2.2.4.2 Electronic Flow Control System-EFCS
The EFCS uses solenoid proportional valves to control the oxygen, air and nitrous oxide flow and
monitor the actual flow, and provides feedback to the flow control unit. After passing three
independent check valves, the gases of the three limbs are mixed at the rear end of the assembly
and output. The O2-N2O ratio is fully controlled by a software algorithm. The system display
outputs the monitored values of parameters such as the limb flow and O2-N2O ratio.
The above figure shows the pneumatic block structure of the EFCS. The three gases, after being
regulated, goes through the proportional valves, through limb sensors, through the gas mixing
chamber, and into the three-way valve. Then, the gas mixture returns to the total flow sensor and
finally output to the mechanical float flow meter.
O2 (from the
O2 supply)
N2O (from
the N2O
supply)
Air (from the
air supply)
From
backup O2
flowmeter
To the
vaporizer
manifold
Balance gas
sensor
O2 sensor
Total flow
sensor
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...