8-28
2. Set the machine following the instructions on the screen, and then select Continue to
perform the manual circuit leak and compliance test. Progress of the manual circuit leak
and compliance test is displayed on the screen. As shown in the figure below, select
Cancel to cancel the ongoing test.
3. After the test is completed, confirm whether the manual circuit leak and compliance test
is passed following the instructions on the screen. If not, deal with it following "Help"
on the screen and the following "Note".