4-9
4.7.2 Checking Needle Valve Correspondence
Perform the following operations in the BFCS state:
Turn the O2 needle valve and verify that the flow on the total flowmeter continuously increases
from about 1 L/min to about 10 L/min. Turn off the O2 needle valve. The flow should decrease to
about 1 L/min (that is, only the basal flow is reserved).
4.7.3 Checking the Needle Valve Sensitive Switch
Press the emergency button. After the BFCS is deployed, click Disable
Backup Flow Control
or
press the emergency button again. A prompt message "
Close manual valves prior to disabling
Backup Flow Control
" is displayed. The prompt message disappears and the BFCS can be
normally disabled.
4.7.4 Basal Flow Test in the BFCS State
1. Connect the O2 pipeline supply or open the backup O2 cylinder.
2. Press the emergency button to deploy the BFCS.
3. Verify that the float of the total flowmeter is in the vicinity of 1 L/min.
The following operations are optional:
1. Set the system to work in ACGO mode.
2. Connect the pipeline from the ACGO to the inlet of the flow test equipment.
3. Verify that the basal O2 flow is 1±0.25 L/min.
NOTE
The gas to be tested must be O2 and the O2 needle valve must be turned off.
4.8 Flow Control System Tests
4.8.1 Without O2 Sensor
NOTE
Turn the flow controls slowly. Do not turn further when the flowmeter indicates
the maximum or minimum flow to avoid damaging the control valve.
WARNING
If N2O is available and flows through the system during this test, use a safe and
approved procedure to collect and remove it.
Incorrect gas mixtures can cause patient injury. If the O2-N2O Link system does
not supply O2 and N2O in the correct proportions, do not use the system.
To do the flow control system tests:
1. Connect the pipeline supplies or slowly open the cylinder valves.
2. Turn all flow controls fully clockwise (minimum flow).
3. Set the system switch to the
position.
4. Connect the AC power source if battery capacity shortage occurs. Do not use the system
if other ventilator failure alarm occurs.
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...