2.2.4 Anesthetic Gas Delivery System
2.2.4.1 System Switch Assembly
There are two kinds of systwm switch, single O2 system switch, and O2 and Air system switch.
The above figure shows the O2 and Air system switch. Supply gases of Air and O2 go into system
switch 22, and then needle valve. For single O2 system switch, only O2 goes into system and then
needle valve. System switch has an electrical outlet which controls the power-on status of the
system. When the system switch is turned on, O2 and Air enter theneedle valve and the system is
powered on simultaneously. The anesthetic ventilator starts to monitor the status of the system.
When the system switch is turned off, O2 and Air cannot enter the needle valve and the system is
powered off.
Air inlet
O2 inlet
O2 outlet
Air outlet
inching
switch
inching
switch