2-31
2.1.6.2 Monitoring Module
Monitoring module monitors pressure and flow of anesthesia machine and breathing system,
controls valves, monitors and collects statuses, reads O2 concentration, monitors pressure and
flow in circuit, and control accuracy accurately. Monitoring module contains VCM and VPM.
VCM monitrs the parameters including inspiratory flow, expiratory flow, interior flow of the
machine, airway pressure, PEEP, O2 concentration, etc, and controls the actions of three-way
valve and exhalation valve (including safety valve, PEEP valve, and inspiration valve).
VPM monitors the signal of the pressure switch of drive gas, ACGO limit switch, O2 pressure
switch at gas source inlet, auto/manual switch, CO2 absorbent canister in-place switch, etc. VPM
also controls the selector valve of drive gas, and controls safety valve with VCM, and send
monitoring information to the main board through series port.
J1, power supply
port and VCM
communication port
J5,VPM
communication port
J7,
communication
port of
calibration
J10,CO2
Absorber
canister
switch signal
J4,
exhalation
valve drive
J8, on/off
signal
J15, control
selector valve
of drive gas
J2,
Exhalation
valve Power
supply
J12, O2
sensor port
J6, three-
way valve
drive
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...