6-1
6
Device Maintenance
WARNING
When it comes to test and maintain the equipment, make sure that the patient is
disconnected from the equipment.
The equipment may have been used on patients carrying infectious diseases. Before
testing or maintaining the equipment, wear sterile rubber gloves to reduce the risk
of being infected.
When the equipment to be maintained contains blood or other secretion, clean,
disinfect and sterilize the equipment by strictly
6.1 Maintenance Overview
Physical check, disposables replacement, and performance check shall be conducted at regular
intervals according to the periods listed in the following table. The factory will bear no liability if
some parts are damaged or lost because the disposables are not replaced within the suggested
period.
6.2 Maintenance Period
Maintenance Operation
After Each
Maintenance
Every 12
Months
Preoperative check:
1. Check the Mechanical Ventilation Mode
2. Breathing System Leak Test in Mechanical
Ventilation Mode
3. Breathing System Leak Test in Manual Ventilation
Mode
4. Check the Sensor Zero Point
5. Check the Flow Sensor Accuracy
6. Check the Pressure Sensor Accuracy
7. Check the Electronic Flowmeter Accuracy
●
System check
●
Disposables replacement
●
Battery maintenance and
replacement
Function test after maintenance
●
Test before operation
●
●
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...