7-15
interface. Press [
ZERO
] on the panel. In the displayed interface, press [
OK
] to complete zeroing.
(2) On the panel of the calibration device VT, press [
MODE
]. Select [
Calibration Mode
] from
the menu, and press [
OK
] on the panel to go to the calibration interface. See the following figure:
If the VT Plus is used as the calibration device, configure the Fluke VT Plus in the following way:
(1) Gas settings: Select [
Main
], and select [
Setting
] > [
ENTER] > [Gas Settings
] >
[
MODIFY
] > [
Gas Type
] > [
O2
] > [
BACK
] > [
BACK
].
(2). Correction Mode: Press the [
Setup]
button, and select [
Setting]
>
[
ENTER]
>[
Correction Mode
]> [BTPS]> [
BACK
]> [
BACK
]
。