8-21
Message
Alarm
Priority
Cause
Solution
C High Imped.
electrode C is too high.
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
C Lead Off
Low
BIS electrode C is
detached from the patient.
1. Check whether BIS electrode C is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
LE High
Imped.
Low
The impedance of BIS
electrode LE is too high.
1. Check whether BIS electrode LE
is in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
LE Lead Off
Low
BIS electrode LE is
detached from the patient.
1. Check whether BIS electrode LE
is in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
LT High
Imped.
Low
The impedance of BIS
electrode LT is too high.
1. Check whether BIS electrode LT
is in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
LT Lead Off
Low
BIS electrode LT is
detached from the patient.
1. Check whether BIS electrode LT
is in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
RE High
Imped.
Low
The impedance of BIS
electrode RE is too high.
1. Check whether BIS electrode RE
is in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
RE Lead Off
Low
BIS electrode RE is
detached from the patient.
1. Check whether BIS electrode RE
is in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
RT High
Imped.
Low
The impedance of BIS
electrode RT is too high.
1. Check whether BIS electrode RT
is in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
RT Lead Off
Low
BIS electrode RT is
detached from the patient.
1. Check whether BIS electrode RT
is in proper contact with the skin.
2. Replace the BIS sensor.
BIS DSC Err
Low
BIS DSC communication
error
4. Replace the BIS module.
BIS DSC Malf Low
The BIS DSC is faulty
and is automatically shut
down.
4. Replace the BIS module.
BIS No Cable
Low
The BIS cable is not
connected.
1. Check the BIS cable.
2. Replace the BIS cable.
BIS No Sensor Low
The BIS sensor is not
connected.
1. Check the BIS sensor.
2. Reinstall the BIS sensor.
BIS Sensor
Too Many
Low
The BIS sensor is
excessively used.
1. Check whether the use times of
the BIS sensor exceeds the limit
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...