9-12
5. Remove the four (4) screws fixing the encoder board and the connected cables to take
out the encoder board.
9.2.13 Remove Display and Touch Screen
1. Remove the display assembly.
2. Unscrew the seven (7) screws fixing the rear cover plate of display.
3. Disconnect the cables connected with the display and the touch screen to take out the
display and the touch screen.
4. Unscrew the four (4) screws fixing the display from a side to remove the display.
5. After the rear cover plate of the display is removed, remove the touch screen from the
front cover of the display.
Touch screen
Display