9-62
WARNING
Sodalime is a caustic substance and is a strong irritant to eyes, skin and respiratory
system. Affected areas should be flushed with water. If irritation continues after
flushing with water, seek medical assistance immediately.
9.4.6 Remove CO2 Bypass Assembly
1. Remove the absorbent canister as per section 9.4.5
.
2. Press inward the fasteners on both sides and the CO2 bypass assembly will drop down
for removal.
9.4.7 Remove Drain Valve
1. Remove the CO2 bypass assembly as per the section
9.4.6
.
2. Lift up and remove the CO2 Absorber Base. Unscrew the screws as shown in the picture
to take out the drain valve. Remove the CO2 Absorber Hose by squeezing the retaining
clips inside the absorber base.