8-49
8.4.3.1 Leakage Test on the O2 Flush Button Assembly
Use the following tools to perform the leakage test on the O2 flush button assembly:
One negative pressure ball
One 3106-06-00 adapter connector
One PU tube (6X100)
Test procedures:
1. Switch off the gas source switch and switch on the O2 flush switch to release the
remaining pressure.
2. Remove the panel of the workbench. Remove the PU tube 52 connecting to the O2 flush
switch assembly and ACGO assembly. Disconnect the ACGO end.
3. Connect the negative pressure ball inlet to the PU tube 52 using the 3106-06-00 adapter
connector. Squeeze the negative pressure ball to exhaust the internal gas.
4. Release the negative pressure ball. If the negative pressure ball is completed expanded
within 30s, the O2 flush switch assembly is damaged.
8.4.3.2 Leak Test for the Auxiliary Oxygen Flowmeter Assembly
Tools:
Negative pressure ball (quantity: 1)
3106-06-00 adapter connector (quantity: 1)
3126-06-00 tube plug (quantity: 1)
PU tube (6 x 100) (quantity: 1)
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...