3
Installation Guide
3.1 Preparation
Before installation, contact the consumer and tell him/her to provide the following materials. The
customer is responsible for supplying these materials. Missing items may result in delays,
incomplete installations, and/or additional service visits.
Gas supply hose connector.
O2 cylinder, N2O cylinder, and Air cylinder in compliance with the configuration
standard.
Anesthetic vaporizer and adapter (if not purchased together with the anesthesia system)
Anesthetic agent
Loose fill or Pre-pak CO2 absorbent
Effective O2, N2O, and Air gas sources (280 to 600kPa (40 to 87psi))
Dropdown hoses for ceiling mounted medical gas equipment, compatible with
quick-disconnect hoses ( if not purchased together with the anesthesia system)
3.2 Assembly
NOTE
After the flow sensors are calibrated, the anesthesia machine must be used together
with the breathing system module. If the breathing system module is removed,
make sure that it is re-installed onto the original anesthesia machine. If other
breathing system module is to be installed, the flow sensor must be re-calibrated
(please refer to 7.3.2Flow calibration (Service))
.
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...