4-1
4
Testing
WARNING
After servicing the equipment or replacing its components, complete all the tests in
this section.
Before doing the tests in this section, completely reassemble the equipment and
refer to 6 Device Maintenance to do necessary calibrations.
4.1 System Inspection
NOTE
Make sure that the breathing circuit is correctly connected and not damaged.
The top shelf weight limit is 30 kg.
WARNING
Do not leave gas cylinder valves open if the pipeline supply is in use. Cylinder
supplies could be depleted, leaving an insufficient reserve supply in case of pipeline
failure.
Make sure that:
1. The equipment is not damaged.
2. All components are correctly attached.
3. The breathing circuit is correctly connected and the breathing tubes are not damaged.
4. The vaporizers are locked in position.
5. The fixing nuts of the negative pressure suction system are fastened.
6. The gas supplies are connected and the pressures are correct.
7. Cylinder valves are closed on models with cylinder supplies.
8. The casters are not loose and the brake (s) is set and prevents movement.
9. Make sure the circuit is locked safely (in the
position).
10. The power cord is correctly connected. The AC mains indicator and the battery indicator
work normally.
11. The anesthesia machine is switched on or off normally.
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...