2-55
2.2.4.5 ACGO Assembly
The above picture shows the ACGO assembly. The ACGO assembly includes flow restrictor 21,
pressure relief valve 31, and pressure relief valve 55. Inputted O2 and fresh gas are mixed and
enter the ACGO. Pressure relief valve 31 at the front restricts the pressure of inputed O2 and also
that of the fresh gas not to exceed 37.9 kPa. Pressure relief valve 55 at the back ensures that the
pressure of the gas outputted to the ACGO does not exceed 12.5 kPa. Flushing O2 and fresh gas
are mixed through the three-way valve and enter the ACGO. The outputs include fresh gas
provided for the breathing system (when ACGO is turned off) and that provided for the patient
(when ACGO is turned on). Pressure relief valve 31 at the front restricts the pressure of flushing
O2 and also that of the fresh gas not to exceed 38 kPa (approximate value). Pressure relief valve
55 at the back ensures that the pressure of the gas outputted to the ACGO does not exceed 12.5
kPa.
2.2.4.6 O2 Flush Button Assembly
O2 inlet
Enter breathing
system
ACGO outlet
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...