4-18
a. Turn the top cover counterclockwise to separate it from the sight glass. Remove the
filter screen.
b. Shake dust and foreign substance from the removed filter screen until satisfactory
clean effect is achieved.
3. The waste gas disposal system is not working or the pump rate is less than the AGSS
normal working flow. Check if the waste gas disposal system reaches the pump rate
range which the AGSS declares, which is 25 to 50 L/min for low flow AGSS and 75 to
105 L/min for high flow AGSS.
4.10.5.2 Check the Transfer Tube and Active Scavenging Tube
1. Disconnect the tubes from other components.
Check the transfer tube and its connectors for damage.
Check if the transparent silicone hose for damage.
2. Check the receiving hose regarding the following three aspects:
a. The receiving hose and its connectors for damage.
b. If the connections between the receiving hose and its connectors are loose.
c. The sealing ring for damage.