3-2
3.2.1 Unpacking and Setup
1. After receiving this device, check the packing box for damage immediately.
a. If there is no damage and the anti-tilt label on the packing box does not turn red, sign the
name and date on the bill of lading or air waybill to indicate that the anesthesia machine
has been received safely.
b. If there is damage or the anti-tilt label on the packing box is not in good condition (red),
describe the damage on the bill of lading or air waybill. Both the carrier and the receiver
must sign his name and date on the bill of lading or air waybill. Save all damaged
delivery package until further information is given by Mindray. The receiver must
contact Mindray Customer Service Department immediately.
NOTE
When unpacking the equipment, keep as much of the plastic covering on the
equipment. When all parts are unpacked, place the packing materials into the
original packing box. Then, place the smaller boxes inside the large box.
2. Cut, remove, and discard the white packing straps on the packing box.
3. Remove the packing box cover and put it on the floor beside the equipment. Later, the
cover can be used as slope for the convenience of pushing this device onto the floor.
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...