7-45
7.3.6 O2 Sensor Calibration
NOTE
The O2 sensor needs to be calibrated when a great deviation occurs in the O2
monitored value or the O2 sensor or VCM is replaced.
Observe whether the O2 sensor displays values on the measure screen before
calibration. If not, check whether the O2 measure switch is turned on, check the
connection cables of the O2 sensor, or replace the O2 sensor until values are
displayed.
The O2 sensor needs to be calibrated at 21% O2 and 100% O2 and corresponding
standard O2 concentration gases are required.
7.3.6.1 21% O2 Calibration
Perform 21% O2 sensor calibration as follows:
1. Select
[
Main
] > [
General
] > [
Calibrate O2 Sensor
], or select [
Main
] > [
System
] >
[
Calibrate
] > [
O2 Sensor
] or [
Main
] > [
Service
] > [
Calibration
] > [
O2 Sensor
] to go
to the screen as shown in the following figure. The [
General
] tab page displays only the
21% O2 sensor calibration option. The [
System
] and [
Maintenance
] tab pages require
the password for 21% and 100% O2 sensor calibration. You can set the machine based
on the instructions displayed on the screen. Select [
Start
] to start calibration.
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...