9-24
9.2.24 Remove Backup Cylinder Bracket
1. Remove the service door and rear cover plate.
2. Remove pipeline and coil connected with the backup cylinder bracket.
3. Unscrew the four (4) screws fixing each cylinder bracket to remove one backup cylinder
bracket.
Refer to the following table for backup cylinder brackets:
P/N Description
Application
115-002250-01 O
2
cylinder bracket assembly
Without electronic pressure
monitoring
115-002251-02 N
2
O cylinder bracket assembly (
Φ
6)
Without electronic pressure
monitoring
115-002728-01 Air cylinder bracket assembly
Without electronic pressure
monitoring
115-056316-00
O
2
cylinder bracket assembly (0627/electronic
pressure monitoring)
With electronic pressure
monitoring
115-056317-00
N
2
O cylinder bracket assembly
(0627/electronic pressure monitoring)
With electronic pressure
monitoring
115-056318-00
Air cylinder bracket assembly (0627/electronic
pressure monitoring)
With electronic pressure
monitoring
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...