7-8
7. The screen shown below is displayed upon a successful flow sensor calibration. Select
[
Done
] to exit the calibration screen.
NOTE
If measurement deviations are not corrected after multiple flow sensor
calibrations, it is recommended that a flow sensor in the circuit be replaced and
then a calibration be performed for the new flow sensor. If the problem persists,
send the machine to the manufacturer for maintenance. After the problem is fixed,
perform the calibration and system tests.