8-91
Failure
Description
Possible Cause
Judgment Method
Recommended
Action
The alarm
Heating
Module Failure
is generated at
startup and the
circuit does not
heat.
The heater is in
poor condition.
Measure the resistances of pins 1 and
2 and pins 3 and 4 of the connector
that connects to the thermistor under
the workbench by using a multimeter.
If the difference between the
measured resistances exceeds 300
ohms, the heater is faulty.
Replace the
heater.
The power board
is in poor
condition.
If the difference between the
measured resistances is small, the
power board is faulty.
Replace the
power board.
During the
switching of
the
bag/mechanical
ventilation
switch, the
ventilation
mode
displayed in
the upper left
of the screen
does not
change.
The material of
the
bag/mechanical
ventilation switch
is in poor
condition.
Power off the anaesthesia machine,
open the cover plate of the
workbench, test pins 1 and 2 of the
interconnection terminal of the
bag/mechanical ventilation switch by
using the conduction position of the
multimeter, and turn the
bag/mechanical ventilation switch. If
the conduction of pins 1 and 2 does
not change, the material of the
bag/mechanical ventilation switch is
in poor condition. If the conduction
changes, proceed with the analysis.
Replace the
bag/mechanical
ventilation
switch.
The cable is not
connected
properly.
Power on the anaesthesia machine,
test the voltage of pins J8-12 of the
VCM by using the multimeter. The
voltage should be 0-0.4 V in bag
ventilation mode and be 3.0-3.6 V in
mechanical ventilation mode. If the
measurement value is beyond the
normal range, the cable is not
connected properly. If the
measurement value is within the
normal range, proceed with the
analysis.
Replace the cable
of the VCM.
The VCM is in
poor condition.
If the fault cannot be identified by
using the preceding two methods, the
VCM is abnormal.
Replace the
VCM.
The
anaesthesia
machine
The cable is not
connected
properly.
Check whether the two connectors of
the display data cable are connected
properly. If so, proceed with the
Connect the cable
properly.
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...