9-11
9.2.11 Remove the Alarm Lamp Board
1. Remove the display assembly.
2. Remove the cables from the alarm lamp board.
3. Remove the two (2) screws on the alarm lamp board to remove the board.
9.2.12 Remove Encoder Board and Encoder
1. Remove the display assembly.
2. Disconnect the cables connecting the encoder assembly.
3. Remove the three (3) screws fixing the encoder assembly.
4. Remove the encoder assembly.