7-26
Failure Description
Possible Cause
Recommended Action
[00002000] is
displayed
The resolution of the expiratory
flow sensor is incorrect.
1. Check the sampling line connection
and tightness.
2. Check the gas supply pressure.
3. Check the configuration of the
calibration device.
4. Troubleshoot the sensor and valve
fault using the valve diagnosis tool.
For details, see section 5.4.
5. Replace the corresponding flow
sensor on the circuit.
6. Replace the VCM.
[00004000] is
displayed
The resolution of the ventilator
flow sensor is incorrect.
1. Check whether the sampling line is
connected correctly.
2. Replace the ventilator flow sensor.
3. Replace the VCM.
[00008000] is
displayed
The valve output flow speed is
insufficient.
1. Check whether the gas supply
volume is sufficient for the whole
calibration process.
2. Check whether the maximum
output flow on the valve is greater
than 90 L/min. If not, replace the
inspiratory valve.
[00010000] is
displayed
The valve resolution is
insufficient.
1. Check whether the gas supply
volume is sufficient for the whole
calibration process.
2. Check whether the calibration
device works properly.
3. Replace the inspiratory valve.
[00020000] is
displayed
The flow change is not
unidirectional.
1. Check whether the tubes are
connected correctly based on the
instructions.
2. Check whether the gas supply
volume is sufficient for the whole
calibration process.
3. Check whether the calibration
device works properly.
[00040000] is
displayed
The communication connection
to the calibration device is
interrupted.
1. Check the communication
connection between the calibration
device and the anesthesia machine, or
reconnect the calibration device and
the anesthesia machine to ensure that
communication connection works
properly. If the communication
connection does not work properly,
replace the communication cable.
2. Check the configuration of the
calibration device. Re-configure the
device if necessary.
[00080000] is
displayed
The system fails to write to the
EEPROM.
1. Perform the calibration again.
2. Replace the VCM.
[00100000] is
displayed
The ACGO switch is in ON
position.
1. Check whether the ACGO switch is
in OFF position.
2. Check the monitoring status of the
ACGO switch on the interface of the
anesthesia machine.
Summary of Contents for WATO EX-55Pro
Page 1: ...WATO EX 55Pro 65Pro Anesthesia Machine Service Manual ...
Page 2: ......
Page 34: ...2 20 South African standard ...
Page 48: ...2 34 VPM ...
Page 142: ...4 42 ...
Page 196: ...6 26 FOR YOUR NOTES ...
Page 356: ...8 102 FOR YOUR NOTES ...
Page 438: ...9 82 FOR YOUR NOTES ...
Page 447: ...10 9 10 3 2 Hardware Box 2 4 3 5 7 8 6 12 9 10 13 14 15 1 11 ...
Page 449: ...10 11 ...
Page 455: ...10 17 10 3 5 WATO EX 55 Pro 65 Pro Cart Assembly ...
Page 457: ...10 19 10 3 6 WATO EX 55 Pro 65 Pro Pre pak Circuit Bracket Assembly ...
Page 462: ...10 24 4 3 1 2 ...
Page 473: ......
Page 474: ...PN 046 009688 00 4 0 ...