ES601-US OPERATING MANUAL
Manual Testing/Chapter 5
#
Page 69
Figure 37
- DUT connection instructions for AAMI safety test.
STOP BEFORE
SUPPLY OFF:
If set to "YES", this control will cause the test to
fully stop
immediately before
a step which disconnects the AC supply from the DUT. ES601-US will prompt
the user to shut down the DUT and will not proceed until the user
acknowledges by pressing the
CONTINUE
button. After
CONTINUE
is
pressed, ES601-US will then disconnect the AC supply from the DUT. This
feature accommodates equipment having special shutdown requirements, such
as diagnostic imaging equipment or equipment incorporating a PC.
BACK:
Returns to the previous menu (leakage test conditions or patient lead setup, as
appropriate).
NEXT:
Displays instructions for connecting the DUT to ES601-US. The instructions
displayed will depend on the safety standard. The example menu shown
following is for the AAMI-ES1 standard.
According to the CONNECTIONS menu, at minimum a AAMI safety test will require the user to connect
the DUT to the test receptacle and the Kelvin Cable of ES601-US to the enclosure of the DUT. The power
switch of the DUT must also be set to "ON" before the test is started. Depending on how the user has
chosen to set up the test via preceding menus, additional connections may be required between AP jacks on
ES601-US and DUT patient leads; between AUX1 of ES601-US and DUT signal I/O; and between AUX2
and exposed ungrounded metal on the DUT.
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...