ES601-US OPERATING MANUAL
Automated Testing/Chapter 6
#
Page 130
6.1.5.10 Checkitem Summary
Op Code
Checkitem Type
Buttons shown on LCD
A
Run AutoSequence
ESCAPE; BACK; SKIP; CONTINUE
C
Tester Remote Control
If no tester response: ESCAPE; BACK; NEXT
D
Timed Delay, in seconds
ESCAPE; BACK; SKIP
E
Data Entry with Pass/Fail
ESCAPE; PASS; FAIL; NOT APPLICABLE;
BACK; SKIP; ENTER TEST DATA
I
Input Requested
QWERTY touch keypad
K
OK Prompt
ESCAPE; BACK; OK
M
Multi-function
If tester control command:
ESCAPE; PASS; FAIL; BACK; SKIP
P
Pass/Fail
ESCAPE; PASS; FAIL; BACK; SKIP;
FAULT CORRECTED
R
Tester Remote Control with
User Data Entry
ESCAPE; PASS; FAIL; NOT APPLICABLE;
BACK; SKIP; ENTER TEST DATA
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...