ES601-US OPERATING MANUAL
ES601pc-US Software/Chapter 8
#
Page 194
8.12.3.2 medTester-to-ES601 Translation Table
The Translation Table of the Options dialog (
Figure 140
) shows how medTester Autosequence checkitems
(designated in the CMMS by "A" checkitems) will be converted by ES601pc-US into corresponding
AutoSequences for the ES601-US analyzer (designated by commands listed at extreme right in the table,
such as "SAF..." or "UT..."). On installation of ES601pc-US, default AutoSequence files for the ES601-US
analyzer are installed and the Translation Table of the CMMS Interface is initialized to suit these default
AutoSequences, as shown in
Figure 140
. If you modify a ES601-US AutoSequence in ES601pc-US, it may
then be necessary to choose a different AutoSequence in the CMMS Translation Table, so that the
ES601-US analyzer operates as expected when running a medTester checklist produced by the CMMS.
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...