ES601-US OPERATING MANUAL
MULTI
GND
WIRE
POINT-
TO-PT
Test Protocol
Test Application
Manual Testing/Chapter 5
#
Page 56
YES
NO
YES
Measure resistance up to 10 times,
between Kelvin Cable and "DUT
CASE" lead. The
OFFSET
calibration
is subtracted from all measurements.
For the first measurement, the Kelvin
Cable should be connected to an
equipotential (grounding) stud located
near the DUT, and the "DUT CASE"
lead connects to the DUT itself. For
subsequent measurements, the Kelvin
Cable may be re-attached at arbitrary
locations on the DUT to confirm
continuity of the ground between the
Kelvin Cable and "DUT CASE" lead.
Ground resistance test of permanently-
installed equipment, followed by
continuity test of grounded metal parts
on the equipment.
Example: test of X-RAY equipment.
YES
YES
YES
Measure resistance between Kelvin
Cable and the test receptacle of
ES601-US
, one time. After this
measurement, the user disconnects
the DUT line cord from
ES601-US
and connects the "DUT CASE" lead
to the DUT.
For measurement #2, 3, 4... , the
resistance between the Kelvin Cable
and "DUT CASE" lead is then
measured, with the
OFFSET
calibration being subtracted to obtain
the results. The Kelvin Cable and test
lead may be moved to arbitrary points
of attachment and up to 9 additional
resistance measurements are acquired.
These additional measurements
confirm continuity of the ground
between the Kelvin Cable and "DUT
CASE" lead.
Ground resistance test of cord-
connected equipment, followed by
continuity test of grounded metal parts
on the equipment.
Example: test of a hospital bed.
Only those combinations shown in the above table are permitted by ES601-US; other combinations of
settings are inhibited.
Referring back to the Sequence Control menu, other buttons provided on the menu are:
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...