ES601-US OPERATING MANUAL
Automated Testing/Chapter 6
#
Page 107
6.1.3 Tester (Test Device) AutoSequences
In addition to Safety and ECG AutoSequences, ES601-US provides for other types of automated tests. On its
left side panel, ES601-US provides six RS232 serial interfaces or "TESTER PORTS", which can be connected
to external measuring instruments or "
Testers
" by means of accessory adapter cables. Tester AutoSequences of
ES601-US remotely control these external Testers via the TESTER PORTS, allowing ES601-US to
automatically execute tests of medical device performance and acquire and store the results.
ES601-US provides Tester AutoSequences for Datrend products such as the
Oxitest Plus 7
pulse oximeter
tester,
Infutest 2000
infusion pump tester, and
Phase 3
defibrillator/pacemaker tester. AutoSequences for
Oxitest, Infutest and Phase 3 can be user-programmed and uploaded to the ES601-US analyzer by means of
corresponding modules of the ES601pc-US software package, which is described in Chapter 8.
Optionally, ES601-US can be interfaced to test instruments from manufacturers other than Datrend. The
Inspections and User-Tester Editor module of ES601pc-US software allows the user to program and upload
AutoSequences for up to five
User-Defined Testers
or analyzers. Compared to setting up Oxitest, Infutest and
Phase 3 AutoSequences with ES601pc-US software, AutoSequence programming for a User-Defined Tester
requires greater knowledge on the part of the user, for example, knowledge of the remote command language of
the Tester and an understanding of ES601-US
Checkitems
and
Checklists
, which are explained later in this chapter.
As illustrative examples, ES601-US is provided with default example AutoSequences installed for Fluke
Biomedical
Impulse 4000
defibrillator/pacemaker tester, as well as for Pronk
SimCube
®NIBP monitor tester.
Accessory cables are available for connecting ES601-US to test equipment from various manufacturers. The
table below lists adaptor cables for specific instruments, as well as the default TESTER PORT which can be
used for the RS232 serial connection. Note the TESTER PORT specified in the table is simply a factory default
setting, and this setting can be modified as desired by the user when a test is started; any of the six TESTER
PORTS of ES601-US can be used to connect to the specified instrument, via the specified accessory cable.
Tester
Application
Default Port
Interface Cable
Datrend O7
Pulse oximeter performance testing
P-1
P/N 7500-436
Datrend Infutest 2000
Infusion pump performance testing
P-2
P/N 7500-437
Datrend Infutest Solo
Infusion pump performance testing
P-4
P/N 7500-435
Datrend Phase 3
Defib/pacemaker performance testing
P-3
P/N 7500-435
Datrend Accusim-BP-DT
NIBP monitor performance testing
P-4
P/N 7500-436
Datrend AMPS-1
Patient monitor performance testing
P-4
P/N 3140-429
(1)
Fluke Biomedical Impulse 4000
Defib/pacemaker performance testing
P-3
P/N 7500-438
Pronk Technologies SimCube®
NIBP monitor performance testing
P-4
P/N 7500-436
Clinical Dynamics AccuPulse
NIBP monitor performance testing
P-4
P/N 7500-436
Note (1): Cable 3140-429 is provided with AMPS-1 as a standard accessory
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...