ES601-US OPERATING MANUAL
Setup/Chapter 3
#
Page 24
3.3 System Settings Menu #2
System Settings Menu #2 shows default
Sequence Controls
for MANUAL MODE tests. Buttons on the left of
this menu pertain to manual leakage and load current tests, and buttons on the right pertain to manual
ground resistance tests.
When a leakage or ground resistance test is selected from the MANUAL MODE menu, the applicable
Sequence Controls from System Settings Menu #2 are recalled and are shown on the display. Before
proceeding with the test, the user may modify the default Sequence Controls to suit the specific needs of the
equipment being tested. After the manual test is completed, the Sequence Controls revert to the default
settings shown on System Settings Menu #2. System Settings Menu #2 should therefore be configured to
suit the user's preference for manual tests, based on the requirements of equipment most often tested in
manual mode.
Figure 6
: System Setting Menu #2
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...