ES601-US OPERATING MANUAL
Setup/Chapter 3
#
Page 21
3
Setup
3.1 Powering Up
When the power to ES601-US is switched on, the Start-Up menu is displayed while internal self-tests are
conducted. The Start-Up menu displays the firmware version and the current time and date. After a few
seconds, the Startup Menu is replaced by the MAIN MENU of ES601-US.
Figure 3
- Startup Menu
Figure 4
- Main Menu
Optionally, the "STARTUP MODE" of ES601-US can be set to enable a test of the AC supply voltage at
power-on. If the AC supply test at startup has been enabled, ES601-US will produce an alarm if a reversed-
supply condition is detected. Note that a reversed AC supply may affect test results, particularly the leakage
measurements. In the event of a reversed-supply alarm, connect to a different wall outlet or investigate the
wiring of the wall outlet in use before proceeding.
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...