ES601-US OPERATING MANUAL
Appendix B
#
Page 212
15 CHECK CAUTION/OPERATION LABELLING
...........................................PASS
16 CHECK PADDLES FOR SIGNS OF DAMAGE/PITTING
...........................................PASS
17 CHECK BATTERY AND REPLACE IF NECESSARY
...........................................SERVICED
18 CHECK POWER-ON SEQUENCE
...........................................PASS
19 CHECK FOR UNUSUAL NOISE OR VIBRATION
...........................................PASS
20 CHECK DISPLAY FOCUS, INTENSITY, ALIGNMENT
...........................................PASS
21 CHECK RECORDER OPERATION
...........................................PASS
22 CLEAN INTERIOR/EXTERIOR AS REQUIRED
...........................................PASS
23 ENTER SOFTWARE REVISION LEVEL:
VERSION 5.10.2.............................OK
24 RUN SELF-TEST, IF EQUIPPED
...........................................PASS
25 TEST ALL VISUAL ALARMS/INDICATORS
...........................................PASS
26 ECG01
EXECUTE ECG AUTOSEQUENCE...................OK
27 UT101
7 DEMO AUTOSEQUENCE..........OK
28 SELECT PULSE OXIMETER ON 7
...........................................OK
29 SET HEART RATE=60BPM
...........................................OK
30 SET PULSE AMPL=100%
...........................................OK
31 SET ARTIFACT=OFF
...........................................OK
32 CHECK SPO2=96-98%; RATE=59-61BPM
SPO2=96% RATE=60BPM - OK...................PASS
33 CHECK SPO2=88-92%; RATE=59-61BPM
SPO2=89% RATE=60BPM - OK...................PASS
34 CHECK SPO2=78-82%; RATE=59-61BPM
SPO2=79% RATE=60BPM - OK...................PASS
35 CHECK SPO2=68-72%; RATE=59-61BPM
SPO2=68% RATE=60BPM - OK...................PASS
36 SET SPO2=93%
...........................................OK
37 CHECK SPO2=92-94%; RATE=119-121BPM
SPO2=92% RATE=119BPM - OK..................PASS
38 CHECK SPO2=92-94%; RATE=178-182BPM
SPO2=93% RATE=179BPM - OK..................PASS
39 SET HEART RATE=60BPM
...........................................OK
40 CHECK SPO2=92-94% (LOW PERFUSION)
SPO2=92%, WEAK PLETH SIGNAL................PASS
41 CHECK LOW SIGNAL ALARM
...........................................PASS
42 SET PULSE AMPL=100%
...........................................OK
43 END OF SEQUENCE
...........................................OK
44 MEASURE & RECORD DEFIB OUTPUT AT 50 J
49.8 JOULES................................PASS
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...